FDA does NOT recommend using soap and water solutions to leak test oxygen cylinders.
According to FDA guidance and industry standards, soap and water solutions used to leak test are not recommended and may even prove dangerous. Soap contains animal fats (lard) which burns easily and explosive-like in oxygen. Synthetic soap such as dish washing soap (Dawn, Joy, etc.) contain petroleum (oil). Some soaps contain both.
These solutions leave residues on the valve that, when they come in contact with high pressure oxygen, can ignite and cause a fire.
The FDA requires 2 leak tests during the fill process, one while the cylinder is filling and another on the cylinder after they have been filled. This test must be done on every cylinder, to ensure that the contents of the cylinder is not leaking out, and that the patient will receive the full contents of the cylinder.
1.Certified Medical Grade Leak Detector
Leak Test Solution - One Gallon #1109-1023-1
Meets MIL-PRF-25567E specifications for oxygen compatibility.
2. Everyday Cleaner
Concentrator Cleaner - One Gallon #1109-5527
Concentrator Cleaner Wipes - 120 Count Wipes #1109-5528
No VOC’s, CFC’s, Chlorinated solvents, phosphates, alcohol, ammonia, acids or abrasives.
3. Tough Dirt and Adhesive
Label Remover 3G - One Gallon #1109-5520
Label Remover 3G 32 oz - 32oz Bottle Label Remover 3G #1109-5521
Label Remover 3G Wipes - 120 Count Wipes #1109-5524
No rinsing needed, non-oily, low odor and toxicity.